OC5B
Performance
● Maximum flowrate recommended: 5L/min
● Oxygen Output Concentration Levels: (reach the stated oxygen concentration level within 30 minutes after turning on the machine): 93%±3% at the flowrate of 0.5~5L/min
● Flowrate range: 0.5~5L/min
● Flowrate change at maximum recommended flowrate with 7 kPa back pressure: <0.5L/min.
● Outlet pressure: 20kPa~50kPa
● Input Power Consumption: 400VA
● Electrical Requirements: AC220V±10%, 50Hz±1Hz
● Operating noise: ≤55dB(A)
● SpO2 Measurement
1) Range: 0%~100%
2) Resolution: 1%
3) Accuracy: ±2% when SpO2 is in range of 70%~100%; unspecified when less than 70% .
● Pulse rate Measurement
1) Range: 30bpm~250bpm
2) Resolution: 1bpm
3) Accuracy: ±2bpm or ±2%, whichever is greater
Performance
● Maximum flowrate recommended: 5L/min
● Oxygen Output Concentration Levels: (reach the stated oxygen concentration level within 30 minutes after turning on the machine): 93%±3% at the flowrate of 0.5~5L/min
● Flowrate range: 0.5~5L/min
● Flowrate change at maximum recommended flowrate with 7 kPa back pressure: <0.5L/min.
● Outlet pressure: 20kPa~50kPa
● Input Power Consumption: 400VA
● Electrical Requirements: AC220V±10%, 50Hz±1Hz
● Operating noise: ≤55dB(A)
● SpO2 Measurement
1) Range: 0%~100%
2) Resolution: 1%
3) Accuracy: ±2% when SpO2 is in range of 70%~100%; unspecified when less than 70% .
● Pulse rate Measurement
1) Range: 30bpm~250bpm
2) Resolution: 1bpm
3) Accuracy: ±2bpm or ±2%, whichever is greater